Product Information

  • The Triplett PCAL300 Loop Process Calibrator is designed for benchtop and field industrial process device testing and calibration.
  • Measures DC voltage and current and sources DC mA current with an accuracy of 0.02%.
  • Provides simultaneous 24V loop power to the transmitter while conducting mA current measurement.
  • Features an adjustable 0-20mA/4-20mA span with percentage readout.
  • Supports an automatic sawtooth or triangular output current step function.
  • Includes an automatic power-off function and a built-in flashlight.
  • Includes test leads, three AA batteries, four spare fuses, and a latch key for opening the battery door.
  • Comes with a 1-year warranty.

PCAL300-NIST comes with Certificate of Traceability to N.I.S.T.
Please add 11 business days lead-time for the NIST calibration process

       

CATEGORY: PROCESS CALIBRTORS

 

  • Precision Calibrator for Sourcing & Measuring DC V/mA
  • Simultaneously measure and source with large dual backlit display
  • High accuracy of ±0.05% DCV and ±0.02% DCmA for measuring and ±0.05% for sourcing
  • Switchable 0 to 20mA and 4 to 20mA ranges with % readout for rapid loop verification
  • Auto sawtooth and triangular current outputs for efficient loop testing
  • Switchable 0 to 20mA and 4 to 20mA ranges with % readout for rapid loop verification
  • 0 to 10VDC output with high resolution for precise calibration
  • Built-in flashlight, auto power off, and low-battery indicator for field use
  • Optional models include a Certificate of Traceability to NIST
  • Includes industrial test leads with clips, spare fuses, latch key, and (3) AA batteries.
  PCAL400  
Specifications Range (Resolution) Accuracy
Measure: DC Voltage 100mV, 1V, 35V (0.001mV, 0.1mV, 1mV) ±(0.05%+0.03mV)
Measure: DC Current 30mA (0.001mA) ±(0.02%+4µA)
Source Range (Resolution) Accuracy
Source: DC Voltage 10V (0.001V) ±(0.05%+2mV)
Source: DC Current 20mA (0.001mA) ±(0.05%+4µA)
Power 3 × AA batteries  
Dimensions 7 × 3.5 × 2" (178 × 89 × 51mm)  
     